#General information ITEM section %ITEM SERIAL NUMBER 20220900200759 Mfr serial number SRN10147-00759 #Test information Test section %TEST TEST DATE (DD/MM/YYYY) 13/11/2000 PROBLEM NO PASSED YES Run number 20220900200759 #Test data Data section %DATA TEMPERATURE (C) 25 I_LEAK150V (microA) 0.14293 I_LEAK350V (microA) 0.1862 Substr Origin 048 Substr Orient 100 Substr R Upper (kOhm.cm) 8 Substr R Lower (kOhm.cm) 4 Thickness (micron) 291 Vdep (V) 70 R Bias Upper (MOhm) 1.54 R Bias Lower (MOhm) 1.37 #Comment section %COMMENT #Defects section %DEFECT #DEFECT NAME Implant-Open Implant-Short Oxide-Punchthrough Pinhole 324 Short Open %WEBLINK #DESCRIPTION URL #IV raw data Raw data section %RAWDATA DATA #IV 10 49.15 20 75.55 30 88.03 40 96.31 50 102.87 60 108.35 70 113.1 80 117.47 90 121.68 100 125.67 110 129.65 120 133.26 130 136.58 140 139.79 150 142.93 160 145.86 170 148.71 180 151.45 190 154.09 200 156.62 210 159.05 220 161.5 230 163.77 240 165.95 250 168.18 260 170.28 270 172.28 280 174.25 290 176.18 300 178.03 310 179.81 320 181.4 330 183 340 184.6 350 186.2 #CV 10 15 O.L. 20 O.L. 25 2643.28 30 2297.94 35 2053.98 40 1872.67 45 1730.56 50 1617.04 55 1525.73 60 1459.45 65 1420.63 70 1403.27 75 1396.57 80 1393.56 85 1391.79 90 1390.58 95 1389.63 100 1388.88 105 1388.29 110 1387.73 115 1387.25 120 1386.87 #End of manufacturer data file