#General information ITEM section %ITEM SERIAL NUMBER 20220900208309 Mfr serial number STN12681-08309 #Test information Test section %TEST TEST DATE (DD/MM/YYYY) 27/02/2002 PROBLEM NO PASSED YES Run number 20220900208309 #Test data Data section %DATA TEMPERATURE (C) 27 I_LEAK150V (microA) 0.0868 I_LEAK350V (microA) 0.13028 Substr Origin 231 Substr Orient 111 Substr R Upper (kOhm.cm) 8 Substr R Lower (kOhm.cm) 4 Thickness (micron) 287 Vdep (V) 80 R Bias Upper (MOhm) 1.42 R Bias Lower (MOhm) 1.26 #Comment section %COMMENT #Defects section %DEFECT #DEFECT NAME Implant-Open Implant-Short Oxide-Punchthrough Pinhole 322 Short 752 Short 753 Short 754 Open %WEBLINK #DESCRIPTION URL #IV raw data Raw data section %RAWDATA DATA #IV 10 26.22 20 36.45 30 42.29 40 47.09 50 51.85 60 56.2 70 60.18 80 64.09 90 67.84 100 71.44 110 74.82 120 78.03 130 81.08 140 83.96 150 86.8 160 89.5 170 92.18 180 94.73 190 97.21 200 99.67 210 102.02 220 104.37 230 106.67 240 108.87 250 111.09 260 113.23 270 115.26 280 117.21 290 119.17 300 121.09 310 122.97 320 124.84 330 126.67 340 128.46 350 130.28 #CV 10 15 O.L. 20 O.L. 25 O.L. 30 2593.88 35 2302.62 40 2087.34 45 1920.54 50 1787.06 55 1678.11 60 1591.01 65 1525.42 70 1480.87 75 1453.48 80 1438.35 85 1430.48 90 1426.26 95 1423.86 100 1422.21 105 1421.03 110 1420.1 115 1419.17 120 1418.53 #End of manufacturer data file