#General information ITEM section %ITEM SERIAL NUMBER 20220900209862 Mfr serial number STN12912-09862 #Test information Test section %TEST TEST DATE (DD/MM/YYYY) 22/05/2002 PROBLEM NO PASSED YES Run number 20220900209862 #Test data Data section %DATA TEMPERATURE (C) 29 I_LEAK150V (microA) 0.15181 I_LEAK350V (microA) 0.2198 Substr Origin 261 Substr Orient 111 Substr R Upper (kOhm.cm) 8 Substr R Lower (kOhm.cm) 4 Thickness (micron) 290 Vdep (V) 70 R Bias Upper (MOhm) 1.64 R Bias Lower (MOhm) 1.46 #Comment section %COMMENT #Defects section %DEFECT #DEFECT NAME Implant-Open Implant-Short Oxide-Punchthrough Pinhole Short Open %WEBLINK #DESCRIPTION URL #IV raw data Raw data section %RAWDATA DATA #IV 10 47.14 20 66.04 30 76.67 40 85.02 50 92.11 60 98.76 70 105.21 80 111.81 90 118.6 100 125.53 110 133.05 120 139.73 130 144.06 140 147.92 150 151.81 160 155.64 170 159.54 180 163.27 190 166.91 200 170.49 210 174.04 220 177.31 230 180.5 240 183.7 250 187 260 190.5 270 193.9 280 197.4 290 200.6 300 204 310 207.3 320 210.5 330 213.7 340 216.8 350 219.8 #CV 10 15 O.L. 20 O.L. 25 2797.27 30 2411.28 35 2141.17 40 1938.88 45 1784.86 50 1663.23 55 1570.66 60 1505.63 65 1464.41 70 1440.54 75 1427.61 80 1421 85 1417.52 90 1415.32 95 1413.8 100 1412.62 105 1411.58 110 1410.69 115 1409.88 120 1409.21 #End of manufacturer data file