#General information ITEM section %ITEM SERIAL NUMBER 20220900210311 Mfr serial number STN12926-10311 #Test information Test section %TEST TEST DATE (DD/MM/YYYY) 17/06/2002 PROBLEM NO PASSED YES Run number 20220900210311 #Test data Data section %DATA TEMPERATURE (C) 27 I_LEAK150V (microA) 0.11212 I_LEAK350V (microA) 0.16941 Substr Origin 284 Substr Orient 111 Substr R Upper (kOhm.cm) 8 Substr R Lower (kOhm.cm) 4 Thickness (micron) 290 Vdep (V) 65 R Bias Upper (MOhm) 1.41 R Bias Lower (MOhm) 1.3 #Comment section %COMMENT #Defects section %DEFECT #DEFECT NAME Implant-Open Implant-Short Oxide-Punchthrough Pinhole Short Open %WEBLINK #DESCRIPTION URL #IV raw data Raw data section %RAWDATA DATA #IV 10 35.47 20 49.38 30 56.81 40 62.52 50 67.36 60 72.31 70 77.78 80 83.1 90 88.03 100 92.59 110 96.83 120 100.85 130 104.82 140 108.54 150 112.12 160 115.61 170 118.95 180 122.31 190 125.46 200 128.61 210 131.72 220 134.68 230 137.6 240 140.45 250 143.3 260 146.12 270 148.84 280 151.52 290 154.16 300 156.8 310 159.4 320 161.93 330 164.44 340 167 350 169.41 #CV 10 15 O.L. 20 O.L. 25 2609.4 30 2249.79 35 1997.82 40 1809.37 45 1668.92 50 1565.39 55 1496.57 60 1457.02 65 1436.27 70 1426.66 75 1421.41 80 1418.89 85 1416.75 90 1415.61 95 1414.1 100 1413.24 105 1412.05 110 1411.42 115 1410.57 120 1410.11 #End of manufacturer data file